Vibration Measurement Channel IKV-1-3-1
It consists of a eddy current transducer DS-1 (DS-2, DS-3) and an AS131 transducer (in the measuring unit), or an AS1312 transducer on a connection cable.
It is possible to use on compressor equipment;
Type of output signal - 4 ... 20 mA;
SPECIFICATIONS
METROLOGICAL PARAMETERS
Range of measurement of the range of vibration displacement, microns
0 ... 125, 0 ... 250 (eddy current transducer DS-1 (M8x1))
0 ... 100, 0 ... 250 (eddy current transducer DS-1 (M10x1))
0 ... 500, 0 ... 1 000 (eddy current transducer DS-2)
0 ... 2 000 (eddy current transducer DS-3)Range of working frequencies, Hz
2 ... 500
Installation clearance (default), mm
1.0 ± 0.2 (eddy current transducer DS-1 (M8x1))
1.5 ± 0.2 (eddy current transducer DS-1 (M10x1))
3 ± 0.2 (eddy current transducer DS-2)
5 ± 0.2 (eddy current transducer DS-3)Electrical insulation resistance between the vibration transformer circuit and the shell, MΩ, not less
40
INTERFACE
Type of output
current loop 4 ... 20 mA
Supply voltage range, V - 10 ... 24
EXPLOSION PROTECTION
View - intrinsically safe circuit
Marking
0ExiaIICT6 (eddy current transducer DS-1, DS-2, DS-3)
0ExiaIICT5 (rev. Block)Connection
via TIK-PLC 112 controller
through the security barrier
CONSTRUCTION PARAMETERS
Execution
IKV-1-3-1 isp. A (eddy current transducer
DS-1 (DS-2, DS-3) version 00 or 01 (with connector), converter
AS131, measuring unit, safety barrier
TIK-BIS-52-7 (TIK-BIS-51-7, TIK-BIS-12-1))
IKV-1-3-1 isp. B (eddy current transducer
DS-1 (DS-2, DS-3) version 02 with connector and converter
AS1312 on cable), measuring unit, barrier
Safety TIK-BIS-52-7 (TIK-BIS-51-7, TIK-BIS-12-1))Overall dimensions, mm
D (6.8) L (50; 70; 90) (eddy current transducer DS-1 (M8x1))
D (8.5) L (50; 70; 90) (eddy current transducer DS-1 (M10x1))
D (14.5) L (50; 70; 90) (eddy current transducer DS-2)
D (22.5) L (90) (eddy current transducer DS-3)
98.5x27.5x59 (AS131 converter)
277x217x84 (resize block)Weight, kg, no more
0.35 (eddy current transducer DS-1)
0.4 (eddy current transducer DS-2)
0.5 (eddy current transducer DS-3)
0.1 (AS131 / AS1312 converter)
2.5 (measure unit)Type of mount
threaded hole M8x1 (eddy current transducer DS-1)
threaded hole M10x1 (eddy current transducer DS-1)
threaded hole M16x1 (eddy current transducer DS-2)
threaded hole M24x1 (eddy current transducer DS-3)Degree of protection
IP67 (eddy current transducer DS-1, DS-2, DS-3)
IP20 (AS131 converter)
IP66 (meas. Block)
OPERATING PARAMETERS
Range of working temperatures, ºС
-40 ... + 135 (eddy current transducer)
-60 ... + 135 (eddy current transducer, climatic version CL)
-40 ... + 70 (AS converter)
-60 ... + 70 (AS converter, CL climate version)
MANUFACTURER RELIABILITY AND WARRANTY PARAMETERS
Mean time between failures, hours, at least 10,000
Warranty period, months 18
Service life, years, not less than 10