CEMS Mcs200Hw Systems
(设备/材料)
描述:
The MCS200HW is a multi-component analytical system for continuous monitoring of up to 10 infrared measuring components in the flue gases of industrial combustion plants. The MCS200HW operates in the hot extraction mode: all parts in contact with the medium, from the gas sampling probe to the cuvette, are heated above the dew point and are thus protected against corrosion.
概观:
The built-in lambda probe additionally measures oxygen. Additionally, the measurement of total organic carbon can be supplemented by the integrated GMS811 FIDORi. Internal zero point monitoring allows you to quickly check measured values without span gases. A web display and predefined task assistants in the software make operation very easy.
Short review
- Measure up to 10 IR components plus O 2 and TOC
- Hot Extraction Measurement Technology
- Gas production with virtually no wear thanks to the ejector pump
- Zero point monitoring with internal adjustable filters
- Modbus ® Certified Digital Interface
- Web server for platform independent device management
- Use of dry test gases for HCl and NH 3
Areas of use
- Emission control of waste incinerators and power plants, as well as co-incineration plants, such as cement plants
- Measurement of nitrogen oxides (NO, NO 2 , N 2 O) in nitric acid plants
- Monitoring SO 3 in coal-fired power plants
The exact device data and product specifications may vary and are independent of the respective application and customer specification.
MCS200HW system
Measured values CH 4 , CO, CO 2 , Corg , HCl, H 2 O, H 2 SO 4 , NH 3 , NO, NO 2 , N 2 O, O 2 , SO 2 , SO 3 Measured parameters CH 4 , CO, CO 2 , Corg , HCl, H 2 O, NH 3 , NO, NO 2 , N 2 O, O 2 , SO 2 Measurement principles Interference filter correlation, gas filter correlation Measuring distance 8.48 m Trial amount 200 l / h ... 400 l / h Measuring ranges CH 4 0 ... 70 ppm / 0 ... 700 ppm CO 0 ... 60 ppm / 0 ... 8.000 ppm CO 2 0 ... 25 Vol .-% / 0 ... 50 Vol .-% C org 0 ... 15 mg / m³ / 0 ... 10.000 mg / m³ Hcl 0 ... 9 ppm / 0 ... 1.840 ppm H 2 O 0 ... 40 Vol .-% H 2 SO 4 0 ... 50 ppm / 0 ... 200 ppm NH 3 0 ... 15 ppm / 0 ... 650 ppm NO 0 ... 110 ppm / 0 ... 1.865 ppm NO 2 0 ... 25 ppm / 0 ... 240 ppm N 2 O 0 ... 50 ppm / 0 ... 1.015 ppm O 2 0 ... 25 Vol .-% SO 2 0 ... 26 ppm / 0 ... 875 ppm SO 3 0 ... 50 ppm / 0 ... 200 ppm Certified measuring ranges CH 4 0 ... 50 mg / m³ / 0 ... 500 mg / m³ CO 0 ... 75 mg / m³ / 0 ... 10.000 mg / m³ CO 2 0 ... 25 Vol .-% C org 0 ... 15 mg / m³ / 0 ... 50 mg / m³ / 0 ... 150 mg / m³ / 0 ... 500 mg / m³ Hcl 0 ... 15 mg / m³ / 0 ... 3.000 mg / m³ H 2 O 0 ... 40 Vol .-% NH 3 0 ... 10 mg / m³ / 0 ... 500 mg / m³ NO 0 ... 150 mg / m³ / 0 ... 2.500 mg / m³ NO 2 0 ... 50 mg / m³ / 0 ... 500 mg / m³ N 2 O 0 ... 100 mg / m³ / 0 ... 2.000 mg / m³ O 2 0 ... 25 Vol .-% SO 2 0 ... 75 mg / m³ / 0 ... 2.500 mg / m³ Setting time ≤ 200 s Accuracy ≤ 2%
Relative to the final value of the measuring rangeSensitivity drift ≤ 3%: in the interval between maintenance activities, relative to the final value of the measuring range Zero point drift <3%: final value of the measuring range between maintenance Base point drift <3%: final value of the measuring range between maintenance Detection limit ≤ 2%: relative to the end value of the measuring range Total Organic Carbon (TOC) Measurement 0.05 mg / m³ Comparative accuracy ≤ 3.3%: relative to the end value of the measuring range O 2 measurement ≤ 0.2 Vol .-% Measurement error ≤ 2%
The final value of the measuring rangeWorking temperature ≤ +1.300 ° C Gas temperature Analytic input ≤ +200 ° C Operating pressure 850 hPa ... 1.100 hPa Gas humidity ≤ 40 Vol .-% Ambient temperature +5 ° C ... +40 ° C Storage temperature –20 ° C ... +70 ° C Environmental pressure 850 hPa ... 1.100 hPa Environmental humidity ≤ 90%
Relative humidity; non-condensingConformity Approval for installations requiring approval
2000/76 / EC (17. BImSchV)
2001/80 / EC (13. BImSchV)
27. BImSchV
EN 15267
EN 14181
electrical safety CE Protection class IP54 Analog outputs 0/4 ... 22 mA, 500 Ω
Quantity depending on system configurationAnalog inputs 0/4 ... 22 mA, 100 Ω
Quantity depends on system configuration; decoupledDigital outputs 48 V AC, 0.5 A, 35 W / 48 V DC, 0.5 A, 24 W
Quantity depends on system configuration; decoupledDigital inputs 3.9 V, 4.5 mA, 0.55 W
Quantity depending on system configurationModbus ✔ Type of integration in the bus TCP RTU RS-485 PROFIBUS DP ✔ Note Option Ethernet ✔ Function Binding to SOPAS ET software or OPC server Indication LCD display Status LEDs “Power”, “Maintenance” and “Fault” Enter Touchpad Control Via LCD or SOPAS ET software, multiple access levels, password protection Menu languages German, English Dimensions (W x H x D) 808 mm x 2.208 mm x 623 mm (For details, see the dimensional drawings) Weight OK. 250 kg Material in contact with the medium Stainless steel 1.4571, PTFE, aluminum, coated Source of power Voltage 115 V / 230 V Frequency 50 Hz / 60 Hz Power consumption Analyzer: ± 1.000 W Sample gas sample line, heated: ± 95 W / m Gas Sampling Unit: ± 450 W Heated probe tube: ± 450 W Supporting materials Instrumentation air (zero gas quality) ≤ 350 l / h
6 ... 7 bar; particle size not more than 1 micron; oil content not more than 0.1 mg / m³; pressure dew point –30 ° CInstrumentation air (air as a working fluid for the ejector) ≤ 1.300 l / h
6 ... 7 bar; particle size not more than 1 micron; oil content not more than 0.1 mg / m³; pressure dew point –30 ° CReference gas ≤ 350 l / h
4 bar maximum; reference gas must meet the requirements of applicable directives and standardsSample gas connections Sample gas inlet Compression fitting for 6 mm pipes Auxiliary gas connections Air as a working fluid for an ejector DN 6/8 Reference gas Compression fitting for 6 mm pipes Flue gas outlet DN 8/10 Corrective Functions Drift correction and optical control functions using an alignment cuvette Control functions Automatic control cycle for zero and base points System components Gas sampling unit
Sample gas line
Analytical System Cabinet
Options GMS811 FIDORi
ZIC